CustusX
18.04-rc6
An IGT application
|
The calibration plugin is a collection of calibration widgets and methods.
Calibrates a tool by sampling it when pointing at a known point on another frame. This calibration changes only the translational part of the calibration matrix.
By using the test button you can test your calibration by pointing at a known reference point.
Calibrates a tool by sampling it when it is inserted into the slot of the custom-made Utrecht-SNF-Calibration-tool_POLARIS frame. This calibration changes both the rotational and translational part of the calibration matrix.
The tool to be calibrated will have its calibration set in such a way that the calibrating and calibrated tools are in the same position.
By using the test button you can test your calibration by pointing at a known reference point.
Sample the active tool tooltip in any coordinate system and get the results written to file.
Calibrate the time shift between the tracking system and the video acquisition source.
Note! This widget must be visible during active acquisition for image and tracking data to be stored.
The shift is not applied in any way, this must be done manually after calibration. Refer to the log folder for the calibration curves.
NB: Previous temporal calibration is applied to probe frames during acqusition.
Manipulate the tool calibration matrix sMt directly, using the matrix manipulation interface.